Psychometric test system for measuring cognitive reaction time

  • For the assessment of human reaction time, a test environment was developed. This system consists of an embedded device with organic light-emitting diode (OLED) displays with push buttons for the combined presentation of visual stimulation and registration of the haptic human reaction. The test leader can define the test sequence with the aid of a graphical user interface (GUI) on a personal computer (PC). The validation of the system was proved by measuring the latency times of the whole system, which are conditioned by the specific hard- and software constellation. Through the investigation of the display’s light radiation by a photodiode and the recorded current consumption, latency times and their variance were specified. In the fastest mode the system can reach an error limit of 60 μs.

Download full text files

Export metadata

Metadaten
Author:Klaus P. KochORCiD, Julien Laigle Hoffmann, Lukas Jung, Eric Auer, Hartmut Schächinger
URN:urn:nbn:de:hbz:tr5-2592
DOI:https://doi.org/10.1515/cdbme-2020-3147
Parent Title (English):Current Directions in Biomedical Engineering
Publisher:De Gruyter
Document Type:Article (specialist journals)
Language:English
Date of OPUS upload:2023/03/02
Date of first Publication:2020/11/26
Publishing University:Hochschule Trier
Release Date:2023/03/02
Tag:OLED latency; cognitive performance assessment; reaction time measurement
GND Keyword:Psychometrie; Reaktionszeit; Messung; OLED; Latenzzeit <Informatik>
Volume:6
Issue:3
Article Number:20203147
Page Number:4
First Page:1
Last Page:4
Departments:FB Technik
Dewey Decimal Classification:1 Philosophie und Psychologie / 15 Psychologie
6 Technik, Medizin, angewandte Wissenschaften / 60 Technik